• Continuously adjustable output voltage
  • Fixed overload, factory set to 120% of rated output current
  • Zero start and external interlock provision
  • Dual-range voltmeter
  • Five-range current meter 0-1.0 μA DC with x1, x10, x100, x1k, and x10k multipliers
  • Five-range megohmmeter 100-1 MΩ with x0.1, x1, x10, x100, and x1k multipliers
  • Ferro-resonant line regulator for output voltage stabilization
  • Guard/ground return circuit
  • Automatic transit protected meters
  • Rugged case with cushion grips
  • Transit protected meters

* Due to the ferro-resonant designed input voltage regulator, if an inverter is used for the voltage source it must produce a “true sine wave” output and not a “modified sine wave” (square wave). This model is input frequency sensitive: either 50Hz or 60Hz but not both. Please Specify when ordering.

Input 120V, 60Hz, 5A or 230V, 50/60Hz, 3A (±1% input voltage regulator)*
Output 0-37.5kVdc, 10mA, Negative polarity. positive ground.
Duty Continuous
Kilovoltmeter 3.5″, scaled 0-15/37.5kVdc, ±2% FS accuracy
Currentmeter 3.5″, scaled 0-1.0µAdc, ±2% FS with multipliers x1, x10, x100, x1k, x10k
Guard/ground load return Yes
Megohmmeter Scaled 100 -1MOhms with multipliers of x0.1, x1, x10, x100, x1k
Controls
Voltage output: Continuously adjustable up to max. voltage
Safety
Interlocks: Zero start and external interface
Protection: Surge and transient
Cable Length
Output cable: 6m shielded with alligator clamp
Ground lead: 6m with alligator clamp
Return lead 6m with alligator clamp
Input line cord 3m
Ground stick 6m lead and clamp included
Size & Weight
Controls: 356mm w x 279mm d x 356mm h, 23kg
HV Tank: Included in controls case